AMMRL: signal to noise problem update

From: Julie Fisher <J.Fisher_at_chem.leeds.ac.uk>
Date: Mon, 14 Dec 2009 14:06:05 +0000

Dear All,
I am sorry it has taken so long to get back to you - we have had long delays getting components - and a flooded lab to deal with!
To re-cap :


We have a Varian Unity Inova 500 and ID and H/F/C probes (both 5mm, PFG).

Following a recent probe switch (which may or may not be related to the

problem) S/N levels for 1H/13C (and 19F) is roughly half what it should be and yet pw90 and line shapes are as they should be! I have switched filters, checked cables etc but have not solved the problem.




Many of you (including the Varian UK guys) suggested checking the pre-amp ( I was able to do this through a loan of a pre-amp from another group at Leeds (thanks Arnout)) and this was indeed the problem. Our electronics people set about checking components and spotted two blown components - the FET (Q1, ATF10135) where there was a short between drain and supply and the bipolar amplifier (U1, MSA-1105). These cost 84p (2 week delivery time from US) and £20 (one week delivery time US) respectively.
All is now well.

The student changing the probe had wired cables incorrectly to the preamp (she is not sure what she did but did have to recable when she noticed s/n was so poor)

Thanks again for taking the time to advise on this.
Regards
Julie
*************************************
Dr. Julie Fisher
Reader in Biological NMR
School of Chemistry
University of Leeds
Leeds LS2 9JT
UK
Tel: +44 (0) 113 343 6577
Fax: +44 (0) 113 343 6565
e-mail: j.fisher_at_chem.leeds.ac.uk
http://www.chem.leeds.ac.uk/People/Fisher.html
******************************************
Received on Mon Dec 14 2009 - 04:06:34 MST

This archive was generated by hypermail 2.4.0 : Fri Jun 16 2023 - 15:23:59 MST