AMMRL: Soft Iron shielding "again"

From: Charles L. Anderson <shiulong_at_mail.uh.edu>
Date: Fri, 5 Dec 2008 16:51:06 -0600

Hi everyone

Some follow-up FYI stuff.

The CRT's are used because they are faster than the plasma's when they view
their 3-D subject material.
The 5th fl lab has been reinforced with rebar and wielded wire fabric (WWF)
to support the NMR's weight. It has passed the vender's vibration
requirements so the weight and vibe's are ok. I will give a follow-up on
the shielding once I have all of the info.

Chuck

We visited this back in 2006, but it's come back again from the 4th floor
occupant below the lab. Any new opinions?

"They are building a NMR lab (44'X36') on the 5th floor of a new building.
This lab will have a shielded 500, shielded 400 and unshielded 300 in it.
The architects are going to shield the room with soft iron. This is to
protect the high resolution CRT's that will be used for 3-D visualization.
Anybody had this done to their lab or any comments on this. I am getting a
new lab with 2 new NMR's and I want to get it right the first time. I
believe the lab has been stressed to dampen vibrations from the surrounding
building. However, I'm worried about this soft iron because I know nothing
about it and it has the word iron in it."

 

Thanks
 

Chuck


Charles L. Anderson, Ph.D.
Manager NMR Facilities Office: 713-743-2728
Department of Chemistry Fax: 713-743-2728
302 Fleming Cell: 281-841-7588
University of Houston
Web:nmr-Central.chem.uh.edu
Houston, TX 77204-5003
 
"Anything one man can imagine others can make true"
                                                          Jules Verne
 
 
Received on Fri Dec 05 2008 - 12:54:43 MST

This archive was generated by hypermail 2.4.0 : Wed Jun 14 2023 - 15:07:29 MST