AMMRL: automation time incorrect in chempack4.0 - summary

Paul Ralifo (ralifo@bu.edu)
Tue, 4 Dec 2007 17:03:31 -0500


I recently posted a problem that I was experiencing with vnmrj_1.1_D with 
RedHat Linux 4.0. I received a lot of help from  Dr. Krish Krishnamurthy (of 
Eli Lilly), the automation time is correct when experiments are submitted.
here is the summary:

The problem I was experiencing  was posted on the varian user library site 
as:
PSG - Sequences may ignore delays in exptime calculation
Bug-ID:   psg.j1111  (++,M)
Summary:
(Re-)Compiling any pulse sequence under RHEL 4 leads to an executable that 
ignores the time of all delays during the calculation of exptime. 
Consequently, all timing for manual and automatic data acquisition is 
massively wrong, rendering StudyQ automation impossible. All macros and 
commands that calculate the experiment time are affected (xm1('time'), time, 
exptime). When actually starting the experiment, the correct remaining time 
is displayed in the hardware bar, and the sequence works and gives proper 
spectra. The pre-compiled pulse sequences from the VnmrJ installation CD-ROM 
work OK, as do sequences that were compiled on a RHEL 3/32 bit Linux PC 
(64-bit RHEL 3.2 was not checked).
Bad versions:   VnmrJ 1.1D / RHEL 4 / MERCURYplus
Reported by:   Bert Heise, Varian (2006-10-25)
Priority rating:   P5 (high, "parked")
Status:   Assigned
Workaround:
Use only the pre-compiled pulse sequences in /vnmr/seqlib or compile the 
desired pulse sequence on a Linux computer equipped with RHEL 3, if 
available.
Severity:   1 (Prevents use of system or software package)
At 03:44 PM 9/6/2007, you wrote:

Krish sent me a pulse sequences which he had compiled using RHEL 3 which I 
replaced what I had in my /vnmr/seqlib and this solved the problem.

Thank you


----- Original Message ----- 
From: "Paul Ralifo" 
To: 
Sent: Wednesday, November 28, 2007 3:49 PM
Subject: AMMRL: automation time incorrect in chempack4.0


>I recently installed chempack 4.0 (2001-11-08) version and the latest 
>patches 1.1DallLNXmpl105. I am noticing that the the autmation time for 
>experiments are a little off. For instance if i submit a COSY experiment, 
>it says 4 mins in the studyQ after I update the studyQ, however the actual 
>time the experiment runs is 20 mins. The same was seen for gHMBC and 
>2dNOESY.
> I am running vnmrj_1.1_D on red hat linux 4.0, mercury plus spectrometer. 
> Has anyone noticed this problem. The solutions I have seen on the bug list 
> from varian says to download the latest chempack 4.0 but I already have 
> that. Any suggestions will be greatly welcome.
> thank you
>
> Paul Ralifo
> ___________,,,,>^._.^<,,,,________________
> Paul Ralifo Ph. D.
> Postdoctoral Scientist
> Center for Chemical Methodology & Library Development
> Boston University
> 24 Cummington Street, Rm. 828
> Boston, MA 02215
> (617) 358-0934
>