strange problem of 15NCSA/15N-H dipolar cross corrleation squence on Bruker

From: Weidong Hu <lakesu_at_yahoo.com>
Date: Thu, 11 Dec 2003 13:30:48 -0800 (PST)

Dear Colleagues,

I implemented the experiment to measure 15N CSA/15N-1H
dipolar cross-correlation rate on our Bruker Avance
system. The experiment was published in JMR 127,
128-133pp (1997) from Dr. Geerten W. Vuister group. I
tested the sequence on a standard 15N labeled protein
with about 80 to 480 scans for reference and
cross-correlated experiments, respectively. The
experiments worked fine. When I tried on our real
system with a large number of scan (1920 scan). The
experiment stopped after a few fids, and the error
message is

“ No points acquired during last scan !! (Digitizer
OK?)”.

This errors happed couple of times, I then asked help
from the Bruker engineers, Borlan Pan and Ron Savage.
They suggested replace the HADC/2 digitizer board and
its power supplier. I did that, and the problem is
still there. There are no such problem with smaller
number of scans (480 or so). I start to suspect
something may not be right in my sequence
implementing. I checked them very carefully, and could
not find anything wrong with it. I am wondering did
anyone implement this experiment (the pulse scheme in
Fig. 1A, JMR, 127, 128-133, 1997) on a Bruker
instrument. If you do, could you kindly send me a copy
so that I can find out whether this problem comes from
my implemented sequence or from hardware.

Thank you very much.


=====
Dr. Weidong Hu
Immunology Division, BRI, City of Hope
1500 Duarte Rd., Duarte, CA 91010
Tel: 626 359-8111, ext: 63416
Fax: 626 301-8186

__________________________________
Do you Yahoo!?
New Yahoo! Photos - easier uploading and sharing.
http://photos.yahoo.com/
Received on Thu Dec 11 2003 - 16:20:12 MST