From: L.S. Mombaswala <laiqsm_at_cc.iitb.ac.in>
Date: Tue, 4 Dec 2001 17:17:20 +0530 (IST)

Dear AMMRLs,

        I have VXR 300 S and giving problem in acquisition and lock
channel. The system is communicating wioh host and able to lock with
difficulty, but when the experiment was run gives no signal /FID. On
investigation found that no output frequency and shims where loaded. The
problem was traced to I/O bus and AP Bus problem in acquisition control
Board and Or CPU I/O port.

        Is any one there can help diagnose the problem to its root? On
running existing diagnostic with switch in position 4 gives acquisition
board error or Output board error. I have both Output Board and
Acquisition board and found to have problem with both the boards.

        I was wondering if some one has extra board (acquisition board
with Varian part Number 990640) and prepared to spare. Or else any clue to
go to component level to find the fault ?


        Thanks.

        L.S.Mombasawala.