Inova 500 low band observe problem diagnosed

From: Dan Borchardt <dan.borchardt_at_ucr.edu>
Date: Fri, 01 Dec 2000 13:56:38 -0800

Hi,

I just wanted to thank everyone who responded to my plea for help. After
making a few extra checks suggested by some of you I was able to confirm
that the problem was indeed the Dual Band Mixer Box.

Later,

Dan

Original Message:

>Hi,
>
>Our Inova 500 has a problem. When trying to directly observe 13C there
>is no signal. There is no problem observing proton. I know it is
pulsing
>properly and that rf is reaching the probe via direct observation on a
>scope and the ability to measure indirectly the pw90 of the carbon
>channel. Additionally the built in tune meter does not register (it
>displays 000) when trying to tune the carbon channel but works fine for

>the high band proton channel. I am suspicious of one of the components
>in the magnet leg perhaps the dual band mixer box or the magnet leg
>driver board. So if anyone has had a similar experience or has any
ideas
>as to what the problem could be I would appreciate hearing from you.
>Thanks in advance.
>
>Dan

--