Hello Sameer,

We purchased the Restek leak detector and have found it to be very useful in
finding leaks in our Helium recovery system. The line is long enough so I
can keep the body far enough away from our unshielded 500 without any interference
or draw hazard. It definitely works with in the 5 gauss radius of this magnet.
Not sure if stronger magnets would be an issue, but I think every lab with a
recovery system should have one these to periodically check for leaks.

Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry




> On Aug 23, 2023, at 1:03 PM, Sameer Al-Abdul-Wahid via groups.io <s.wahid=uoguelph.ca_at_groups.io> wrote:
>
> Dear Spinlanders,
>
> Has anyone had successes (or failures) using one of the handheld helium thermal
> conductivity detectors designed for gas chromatography? They typically detect
> down to 1E-5 cc/sec and have a ~1 m long detection hose, allowing us to place
> the unit outside the 5 G line.
>
> Some examples:
> https://www.restek.com/ca/p/28500
> https://www.gow-mac.com/index.php/leak-detector/
> https://www.glsciences.com/product/plumbing_accessories/gas_leak_product/01585.html
>
> We're hoping to find a low-cost alternative to the mass-spec helium detecor on
> campus (which takes ~2 hours to setup), and the MRI-friendly options (e.g., Gascheck
> Tesla or Panther Pro, about 3-4x the cost of the above).
>
> I will post the usual anonymized summary!
>
> Thanks,
> Sameer
> --
> Sameer Al-Abdul-Wahid, Ph.D. | Manager
> Nuclear Magnetic Resonance Centre | University of Guelph
> 50 Stone Rd E | Guelph, ON | N1G 2W1 | CANADA
> 519-824-4120 Ext. 58914 | s.wahid_at_uoguelph.ca
>
>
> http://nmr.uoguelph.ca/
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#476): https://ammrl.groups.io/g/main/message/476
Mute This Topic: https://groups.io/mt/100919310/7559972
Mute #heliumrecovery:https://ammrl.groups.io/g/main/mutehashtag/heliumrecov=
ery
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 23 2023 - 11:22:42 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 25 2023 - 14:43:56 MST