Re: AMMRL: Bruker IconNMR SMTP?

From: Anklin, Clemens <Clemens.Anklin_at_bruker.com>
Date: Mon, 15 Aug 2016 18:46:03 +0000

Hi Bob

how are you.

Sorry to be the bearer of bad news. Email notification with SSL is only possible with ICONNMR5 that comes with Topspin 3.5

Topspin 3.2 does not support it yet, neither did Topspin 2.1.

In Topspin 3.5 you will find the following fields under Options in Iconnmr configuration:

[cid:part1.9F114FB7.F3E276CC_at_bruker.com]

Regards

Clemens


On 8/15/2016 12:48 PM, Robert Hanson wrote:
I have to use smtp.gmail.com<http://smtp.gmail.com>, which uses port 587 and SSL (secure connectivity). I am using TopSpin 2.1, which uses TCL/TK 1.8, so maybe I'm just behind.

Q: Has anyone had any success with IconNMR email notification?

Q: With port 587 and SSL?

Q: With TopSpin 2.1? or a later version of TopSpin?

Bob Hanson

--
Robert M. Hanson
Larson-Anderson Professor of Chemistry
St. Olaf College
Northfield, MN
http://www.stolaf.edu/people/hansonr


If nature does not answer first what we want,
it is better to take what answer we get.

-- Josiah Willard Gibbs, Lecture XXX, Monday, February 5, 1900



--
==============================================================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive Phone:978 667 9580 ext 5144
Billerica MA 01821 skype_ID: canklin
web: www.bruker.com<http://www.bruker.com> e-mail: clemens.anklin_at_bruker.com<mailto:clemens.anklin_at_bruker.com>
==============================================================