AMMRL: Summary: two samples in a magnet

From: Jerry Hu <jghu_at_mrl.ucsb.edu>
Date: Thu, 5 Aug 2010 13:54:20 -0700

Dear Ammrlers,

 

First of all, I would like to thank everybody who has responded to my
inquiry. I learned from each of the replies, which are overwhelmingly
informative!

 

Here is the summary of the replies:

****** Original Question: ******

I would like to hear from you how to deal with and prevent from happening
the accidental presence of two samples in an NMR magnet for liquids.

 

******Answers: ******

Preventive measures:

1. Cheapest: Keep one spinner available at each spectrometer;

2. Most expensive: Get a sample changer;

3. Good education: Reiterate the correct procedures, emphasize the
importance, and stress the consequences;

4. Cap the magnet.

 

Solutions to the problem:

1. Eject: To eject the two samples, need someone to catch the samples at the
top of the magnet because the lower sample will push the 2nd sample out. It
might be helpful holding a hand over the top of the bore and moving it off
quickly to "pop" the samples out, or gradually increasing the eject
pressure.

2. Push from the bottom: first remove the probe, get a second person to
catch the samples from the top of magnet, and then use an appropriate stick
(e.g. helium thumper stick) to push samples up and out;

3. Remove the upper shim stack and then samples;

4. Fish the samples out from the top of magnet.

 

Record for Number of Samples in a Magnet at the Same Time: 4

 

Odds: roughly once every 10 years. Cross fingers, knock on wood if it has
not happened yet.

 

****** Murphy's Law ******

It is an experience common to all men to find that, on any special occasion,
such as the production of a magical effect for the first time in public,
everything that can go wrong will go wrong. Whether we must attribute this
to the malignity of matter or to the total depravity of inanimate things,
whether the exciting cause is hurry, worry, or what not, the fact remains.

 

Cheers,

Jerry
----------------------------------------------------------------------------
--------------------
Jerry Hu, Ph.D. Email:
jghu_at_mrl.ucsb.edu
Project Scientist Tel:
(805)893-7914
Materials Research Lab, UCSB Fax: (805)893-7914
Santa Barbara, CA 93106

 

**************************************

Origina Scripts: (In the order they were received)

#1.

     This doesn't happen to us often, but occasionally someone is oblivious
enough to do it. I have found that usually I can get both samples to eject
without a great deal of problem. Sometimes, I have had to resort to holding
my hand over the top of the bore and moving it off quickly to "pop" the
samples out. That almost always has worked. If not, the only solution is
to remove the probe and drop the shims. That always works.

 

#2. After seeing this happen multiple times at another facility, I've dealt
with the prevention side of that issue here by keeping a single spinner
turbine at each spectrometer console. Although it wouldn't be inconceivable
for someone to walk over to the other instrument and grab the other spinner,
it has not happened so far in over 9 years. Knock on wood!

 

#3. We mostly solved that problem by only providing one spinner per
instrument.

Only very few users put samples in the magnet without a spinner, and usually
only once.

 

#4. Simple solution is to place one and only one spinner at the NMR
workstation. Then users are less likely to walk away with the spinner. If
there is a problem with the spinner, they should report it promptly and you
can swap out the spinner.

 

Also, during the initial new-user training session on our Bruker
spectrometers, I always stress to look at the LED lights on the BSMS
keyboard that indicate if sample is down or missing. I'm sure most users
don't include this point as part of their protocol, but they should (they
have to press lift anyways!).

 

#5. We have a series of small wooden rods that we insert from the bottom
(taping them together as they go in). Eventually they push the samples out
the top. First remove the probehead and insure that the samples are broken,
then insert first wooden rod, tape on the next one, insert next rod, etc.

 

#6. We only let our users have one spinner. We hide the spares.

 We still had a user drop the sample down the bore STILL IN THE BRUKER DEPTH
GAUGE so we put tie-straps on the depth gauge.

 

#7. Owww! Yes, this happened in our lab once. The cause was a slow eject
speed (large 800 MHz magnet, sample had a ways to travel to get to the top);
one of my users got a little impatient, put the second sample in the upper
barrel and hit "insert". Needless to say, we had a "Come to Jesus" meeting
after that to discuss this mishap. I have users put a standard sample in
the magnet and LOCK IT so that there's a non-zero reading on the lock
display; that way, the next user who walks up to the magnet can see that
there's a sample in it. I think the other lab on campus (over in Chem.
Dept.) has users leave the magnet empty and put a cap on the upper barrel to
keep dust out. I was lucky; when I removed the upper barrel, the bottom
part of the lower sample was still intact, so no glass got in the probe.
The top of the bottom sample and the bottom of the top sample weren't so
fortunate. A magnet engineer from Varian walked me thru the procedure over
the phone; the upper barrel on our 800 is a bit non-standard. I might be
able to dig up his name and e-mail and/or phone # if you have a Varian
system.

 

#8. I make sure that there is only one sample spinner available.

 

#9. We had the same problem in the past. I do not know how to prevent it,
except telling the users to eject the sample before they place one. To
eject the two samples, I slowly increased the air pressure for the ejection.
You will need someone to catch the samples at the top of the magnet because
the lower sample will push the 2nd sample out. Hope it helps.

 

#10. I always tell my users the following, "if you don't see the spinner
with a sample in it floating on the top of the magnet and don't hear the
rush of air, don't put a sample in the magnet." I then add, if in doubt,
ask. So far, crossing fingers now that I have mentioned it, no one has put
two samples in the NMR. I will add yet!

 

#11. After carefully lowering and removing the probe, I remove the upper
bore tube and invert it to remove the contents.

Then I beat the devil out of user #2 who put a sample in without ejecting.

 

#12. One way to avoid it is to use ICONNMR and a BACS sample changer, but
since you are in academia, I kind of doubt my suggestion is going to be of
any use to you. But ICONNMR on a BACS protects, and checks that the upper
stack (bore) is empty before inserting the next sample, so it would address
the issue (when operated in 'automation' that is). And I don't even get a
commission from Bruker. ;- )

 

#13. Ha! We've been there. It was like giving birth to twins when you are
only expecting one baby.

Here is the re-constructed scenario on our Varian Inova-300:

1) mystery user #1 did not correctly adjust tube depth in spinner... it was
too far down when replacing the CDCl3 standard in the magnet.

2) NMR tube rested on bottom of probe cavity (miraculously not breaking!)
and stoppered the VT air inlet. This created enough back pressure to pop
apart the black rubber boot on the VT heat exchanger.

3) mystery user does not notice anything wrong and leaves

4) 2nd user arrives and tries to eject standard... nothing comes out because
there is not enough lift due to leak back out disconnected VT fitting.

5) 2nd user assumes there is no sample in magnet and finds another spinner
in cabinet. Puts in sample, in DMSO-d6, this time.

6) locks and shims. Everything looks OK, but spectrum is CDCl3 spectrum,
not DMSO! User #2 calls system manager (me). "The instrument is not

working: I'm getting a chloroform spectrum and my sample was in DMSO."

7) Indeed the spectrometer seems to be working fine. Lock & shim are fine,
but Z0 corresponds to CDCl3, not DMSO, and the spectrum is of chlorform.

8) I eject sample. It lifts up without a problem: indeed, it's the users
sample, not the CDCl3 standard. As I am standing there scratching my head, I
notice that VT air line is disconnected. (VT regulation was turned off).

I reconnect, and Voila! a second sample pops up... the CDCl3 standard!

Miraculously nothing was broken.

There are no bounds to the creativity of user stupidity, but I implemented
the following policies and have not yet had a repeat incidence:

1) all spare spinner turbines are kept under lock-and-key.

    (this would be difficult if you have a sample changer)

2) extensive education on the importance of correctly setting the tube depth
EVERY TIME.

FYI, this is the same lab where someone put a small magnetic stirring bar in
an NMR tube.

 

#14. I only leave one spinner turbine on the instrument so that they have to
remove the first before they can insert the second.

 

#15. We have a policy of having a "dummy sample" (tube of D2O, although it
could probably be water or anything) and a strict policy that a user must
trade their sample with the dummy sample on top of the magnet before and
after they use the spectrometer. I instruct them that if no dummy sample
appears when the eject air is turned on, they must assume that there is a
sample stuck in the spectrometer and get help.

This policy solves two problems -- it is unlikely for a person to try to put
two samples into the magnet, and it is also unlikely that they will forget
to turn on the eject air before inserting their sample.

We also put a spacer around the depth gauge (we used a large rubber stopper
with a big hole in it) to prevent users from trying to drop it into the
magnet.

I also used to restrict the number of spinners in the room to one per
magnet, but now we have a sample changer with many spinners, so this policy
no longer works.

 

#16. Don't do any special to prevent. We get these not too infrequently (at
least a couple times a month). When the two samples/spin collars won't just
come back up under pneumatic pressure, we'll usually drop the probe out, and
use a long 1/4" diameter "thumper" rod to push the assembly out the top of
the magnet; take two people. Often will put a bit of tygon tubing on the
end of the rod, to soften the end that's hitting the bottom tube. Usually
they come out ok.

 

#17. Just leave only one Spinner in the NMR room

 

#18. Sample changers help a lot, but even they don't solve the problem
entirely, sometimes due to equipment failure, and sometimes due to gross
negligence.

 

#19. I suggest only making one spinner turbine available. If there is only
one, then only one can go in. In order to have 2 in there, the 2nd one must
have been dropped in without the Eject air on. We also train our users to
be aware of the air cushion under their sample. If there is no air pushing
up, then they don't put the sample in.

 

#20. You can get them out by removing the upper stack (two thumbscrews, like
a probe, on a Varian) and tipping them out into waiting hands.

You avoid it by educating people. We've tried just leaving one spinner out
(so they have to take the sample out of the magnet to put the next sample
in). But the two sample problem was rare enough that we generally leave two
spinners out. It shouldn't happen because they have to turn the air on to
put the new sample in. It's not hard to be annoyed and martyred enough
while you make the guy stand there while you retrieve his sample, that no
one ever does it twice.

 

#21. There will always be idiots, and even the non-idiots do stupid things
from time to time because they work in their sleep. I have had many
instances of a second sample put in the magnet (and every other imaginable
and non-imaginable mishap) and it's never done any damage.

One sure way to prevent it is to lock up all spinners but one.

 

#22. I always eject the old sample before I insert the new one:) Just
kidding!

 I assume you are not talking about during automation.

 The best way is to have only one spinner turbine. Then you must eject the
old sample before inserting the new.

 

#23. To remove two samples I would use a heium thumper with a tape loop on
the end to hook the top sample and remove it although I have never had to
remove two samples this should work. To prevent it only put one spinner out.

 

#24. I have dummy samples which are loaded into instruments between users.

These are just short lengths of 5mm dia. nylon rod in spinners.

These means that users are always ejecting something before loading a
sample, either a dummy or the last person's sample.

This seems to work well, though I have had a couple of instances of samples
dropped onto the dummy. In one case the sample survived.

 

#25. I am curious to know how that happened... Do you have an autosampler or
is this just someone dropping a tube and spinner without starting the "eject
process"... I always make sure the students pay attention to the sound of
the air flow during inject and eject.

But a serious conversation with the user might be necessary... ;-)

 

#26. We fine the group responsible for the downtime and our time to remove
probes and shims. In my 30 + years at PENN this event has only occured 3
times twice by the same group and person.

 

#27. Just give them only one spinner.

 

#28. Why are you asking? Did this happen to you or are your trying to
prevent a possible occurrence?
Without an autosampler: How in the world can that happen unless you just
drop a sample into the magnet without any ejection air? That would be more
than dumb.
With Autosampler: We have autosamplers on Varian NMRs and only in the
weirdest of conditions can this happen because there are so many safeguards
built in (computer knows there is a sample in the magnet via lock signal, so
does the robot software, etc.). I have heard it happened once, (I wasn't
there) and I believe that a power failure caused the computer to lose the
memory that a sample was in the magnet and the robot did not retrieve the
sample but put the "next" sample into the magnet. In this scenario the only
thing to do is have a plan that ensures that the computer/robot are tested
for automation after such a power failure.
What is more likely to happen is that a chemist puts a sample in the tray of
the autosampler in the position of the sample currently in the magnet.
Robot picks up sample and smashes it into the sample occupying its space
jamming the robot in the process. Not a fun experience. However, the new
robot system on Varian of 400MRs checks that tray location (optically) to be
empty before ejection (if the tray doors have been opened). The older SMS
autosampler cannot do this, but it can be trained to put the robot hand over
that tray location after putting the sample in the magnet, making it
difficult to put a sample in that tray position. If a chemist ignored the
hand and put a sample in that spot i would give them a painful lecture, tell
their supervisor, and require they use the buddy system for several months
to ensure they got the message (just kidding, its never happened, knock on
wood).
Old SMS systems that are not trained to put the hand over the magnet sample
location can numbered sample turbines in EVERY tray location and chemists
must remove a turbine to put their sample tube in and then put it in its
numbered location.

 

#29. For the spectrometers where this is a concern, we only have one spinner
in the room, and when users have finished running, they always leave a
'standard' sample in the magnet. That way, the next user has to eject the
sample to retrieve the spinner. Users who have to use a different spinner
(ceramic or smaller size) have to collect it from a cupboard in another
room, and leave the main Teflon spinner in that cupboard until they are
finished. They are strictly instructed NOT to leave spare spinners lying
around.

This has been very successful in preventing accidental doubling up, however
hasn't eliminated the rare, but ever potential, occurrence of putting the
sample AND depth gauge into the magnet, or the even more difficult to
prevent, insertion of the spinner without the sample! To prevent the former,
we have machined specific depth gauges from plastic, which cannot have their
depth adjusted, and attached a permanent collar to prevent the depth gauge
from fitting into the magnet bore. This became necessary several years ago
following the third such incident in about 18 months. Another option here
would be to tie the gauge to the bench. I don't know how you could prevent
the latter, assuming suitable training has been provided beforehand.

In most cases, retrieval of 'stuck' samples is facilitated using a long,
lightweight Perspex rod fitted on the end with a fresh piece of good quality
double-sided tape (the type for hanging clothing hooks etc to walls). With
the sample lift gas on, the Perspex rod is CAREFULLY inserted into the bore
until it just touches the sample. It is sticky enough to maintain the weight
of a normal sample, especially when being buoyed by the eject gas, and has
worked for removing two samples. Easy does it as you bring it up. I do not
allow any person to do this, other than myself and one other competent
co-worker.

 

#30. hit the lift as usual and catch the first one as it pops out. You might
have to increase the lift pressure through BSMS service, but I have solved
that problem this way on a few occasions

 

#31. We used a copper wire noose on a fiberglass rod to retrieve the top
sample in these instances. As far as prevention we just put emphasis during
user training on sample cleaning/prep/inserting in a specific routine so it
becomes habit.

 

#32. Only one spinner for one magnet.

 

#33. I leave only one spinner with each instrument. A sample is always in
the magnet or being changed. I have had only a couple of problems with
this, mostly the result of cheap NMR tubes.

 

#34. Take out the upper barrel carefully. If Varian system, all of them
should be in the upper barrel.

 

#35. We always have a sample in the magnet. Users do not approach the
magnet to insert a sample without having hit eject first. We occasionally
have someone who drops a sample in but it is a very rare occurrence.
Accidents happen.

I forgot to say that Brandeis 800 users are directed to put on the cap
supplied by Bruker when they remove a sample and do not put another sample
in.

 

#36. in my opinion it can't really be prevented (I'm doing NMR since 20
years and it happend even to me).

Our solution in his case (or if the tube doesn't lift up):

remove the probe head, take a long, but not very flexible plastic tube with
a diameter, slightly smaller than the NMR spinner.

Then try slowly to push up the spinner from below. I 2nd guy should pick up
the samples on the top of the magnet.

We mounted a short piece of such a tube on a flexibel plastic rod to hit the
hole of the ring, where the spinner is sitting, much easier.

On the end of the rod we mounted a long tube to push all to the top.

 

#37. How to prevent: Good training of users. No sample should be inserted
before the gas flow is audible and you can feel it on your fingers. A good
way to make this a habit is to have a sample in the magnet at all times. So
your SOP could just read like .Before a sample is inserted, turn on the
lift, wait until the current sample comes out, remove it and set the new
sample on the air flow. or something.

How to deal with two samples in a magnet? The lift won't work then, so you
have to get them from the bottom, removing the current probe. Which anyway
maybe necessary because it's very likely that one or both of the sample
tubes will be broken, which requires the cleaning of the probe anyway. Just
be very careful to prevent further damage!

 

#38. This particular behavior engineering is greatly helped by rigorously
keeping only one spinner collar available on each system. The users then
must use this one because there is no other. I keep a CDCl3 blank in each
magnet, spinning, thus the students are obliged to eject that tube, insert
their own, then reverse the process when finished. This practice is not
compatible with sample changers, unfortunately.

 

#39. We have two instruments in the same room for 15 years.

We keep only one spinner for the magnet in the area and protocol is to
always have a sample in the magnet.

("Dummy" sample is CDCl3 in quality NMR tube with Teflon cap.)

Never had two samples in the magnet all this time. Other interesting things
happened, but not that.

Now I also have a second location this time with 3 magnets close together -
the wonders of shielding.

Middle one has a sample changer, magnets either side are manual insert.

No problems yet, so I think the protocol with "sample always in magnet"

might be what is helping.

Otherwise if I had to face that sort of problem, I would do something to the
spinners to distinguish them - paint with a sharpie for example - so magnet
1 takes the pink spinner and magnet 2 takes the spinner that is purple and
green dots.

And keep extra spinners put away in a safe place.

Having a magnet with a sample changer worried me for just that reason (all
those spinners!) but somehow all is okay.

 

#40. For our hands-on instruments, we only have one spinner per
instrument--so there's no option of putting another spinner in it.

For our autosamplers, we've had a few times where a sample was broken
inside, but the robot didn't know that and continued to insert and eject
samples. We notice that all the spectra are of the same broken sample, have
the same level of impurity peaks and the peaks start to broaden out as the
solvent evaporates. So far the probe has survived all of this.

 

#41. Just two? The record for our facility if four samples stacked in one
magnet. How did that happen? I don't exactly know because it happened before
I got here, but I suspect one person double-stacked, got confused, and
walked away. Then the next person didn't pay much attention, and inserted
another on top, then another person did the same. Yet someone here at the
time told me it was all one person and one event. I'm not sure what such a
person would be thinking after serial insertions. I've also observed (well,
found out about shortly after the event) people putting tubes in the magnet
with no spinner, putting spinners in with no tubes, putting a sample with a
Bruker spinner in a Varian spectrometer (the B spinners are smaller, and
will go in but won't eject), and people inserting samples with no eject air
on.

 

We've changed the training procedures. We have an old upper barrel, and the
trainees hold on to it while I drop a spinner down it so they can feel the
force of impact as the spinner hits. In the process, I say, "Imagine that
you've spent the last 6 months synthesizing your final product and
everything you have is in this fragile tube .... <bam!> ". I also tell them
the story of the guy who stacked up four samples and make sure they all make
a "what was that guy thinking face" to make sure they know such a thing is
bad. Haven't had any problems from the newly trained people yet.

 

The key procedure all of our users know about is keeping a "lock" (CDCl3)
standard in the probe whenever the spectrometer isn't in use, and making
sure the system is locked. (VNMRJ's "find z0" button has been important for
making it easy to comply with this procedure.) When people walk up to the
instrument, they should see a lock signal on the Christmas tree. If they hit
"eject" and nothing comes out, but they had seen a lock signal beforehand,
they can be certain there's still a sample inside, and they (should) know
enough to not insert their sample and to come get us to check it out. If,
however, the didn't see that the sample was locked, they may be tempted to
assume there was no sample in the magnet and put their own sample in anyway.
I just tell everyone, "If you click eject and nothing comes out, come get
one of us to check it out."

 

 

 