RE: AMMRL: Acquistion Bus Trap Error on Varian Inova #2

From: Dickinson, Thomas <thomas_dickinson_at_merck.com>
Date: Wed, 29 Aug 2007 08:55:31 -0400

Hello,
I have found that the DSP mezzanine can be the root cause for many NP
issues.
Try turning DSP off (dsp='n')and running the experiment again.
Secondly, try swapping the small CTC cable connecting the ADC/DSP to the
Acq. Cont Bd.
Good Luck,
Tom
  

-----Original Message-----
From: rduchesn_at_varianinc.com [mailto:rduchesn_at_varianinc.com] On Behalf
Of Rob Duchesne
Sent: Monday, August 27, 2007 9:55 PM
To: Ivan Keresztes; ammrl_at_ammrl.org
Cc: seamus_at_rice.edu
Subject: Re: AMMRL: Acquistion Bus Trap Error on Varian Inova #2

I would look at the 80 MHz clock signal feeding the digital
card cage. If that is starting to drop out it could cause
some flaky issues with the acquisition. Also, if that is
good look at the 5 VDC power supply for the digital card
cage. The tolerances for the DCC are small and AC ripple can
also wreak havoc.

Good Luck

Rob Duchesne
NMR & MRI Systems
Varian, Inc.

----- Original Message Follows -----
From: Ivan Keresztes <ik54_at_cornell.edu>
To: ammrl_at_ammrl.org
Cc: seamus_at_rice.edu
Subject: AMMRL: Acquistion Bus Trap Error on Varian Inova #2
Date: Mon, 27 Aug 2007 14:38:49 -0400

> Dear All,
>
> The problem described very nicely by Sean in his earlier
> post with the same subject line sounds *exactly* the same
> as the one we have been periodically struggling with over
> the last few years. In our case it occurs on an Inova
> 600. Upgrading to Vnmrj 2.1 changed the error message to
> "Inconsistent DTM error interrupt and DTM status" with an
> occasional "Number of points acquired is [number of
> points] less than np".
>
> I have swapped DTM and DAC boards between the 600 and our
> Inova 400 that is rarely used for long term acquisitions.
> Neither of these swaps fixed the problem. The only
> correlation of any kind may be with number of acquired
> points in t2. The higher the np the more likely that the
> acquisition will fail. Another observation is that any
> particular experiment can be predisposed to failing. We
> had a particular HMBCAD (chempack 4) fail five times in a
> row while other HMBCAD's finished just fine. The problem
> is not restricted to H/X experiments. A number of
> dqcosy's and such failed as well.
>
> I am guessing Sean and I are not the only one to have seen
> this problem and hopeful someone out there has a solution
> for this.
>
> Thanks
>
> Ivan
>
> --
> Ivan Keresztes, Ph.D.
> Director, NMR Facility
> Department of Chemistry and Chemical Biology, Cornell
> University Baker Laboratory, Ithaca, NY 14853.
> phone: (607) 255 0709, fax: (607) 255 4137, email:
> ik54_at_cornell.edu
>

Rob Duchesne
Northeast Region Field Representative
Varian Instruments, Inc.
Office: 650-424-5020 ext 7774
Cell: 978-257-7660




------------------------------------------------------------------------------
Notice: This e-mail message, together with any attachments, contains
information of Merck & Co., Inc. (One Merck Drive, Whitehouse Station,
New Jersey, USA 08889), and/or its affiliates (which may be known
outside the United States as Merck Frosst, Merck Sharp & Dohme or MSD
and in Japan, as Banyu - direct contact information for affiliates is
available at http://www.merck.com/contact/contacts.html) that may be
confidential, proprietary copyrighted and/or legally privileged. It is
intended solely for the use of the individual or entity named on this
message. If you are not the intended recipient, and have received this
message in error, please notify us immediately by reply e-mail and then
delete it from your system.

------------------------------------------------------------------------------
Received on Wed Aug 29 2007 - 12:27:08 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 11 2023 - 16:29:11 MST