AMMRL: TOPSPIN 2.0 Errors

From: Alam, Todd <tmalam_at_sandia.gov>
Date: Thu, 16 Aug 2007 09:50:06 -0600

Dear AMMRL members,


We have an issue here that we have not been able to resolve. This
problem appeared after our recent upgrade to TOPSPIN 2.0 PL5, from
XWINNMR.


During GS, multizg and 2D experiments we will get the following error
message (it never occurs during an individual experiment, only between
experiments). We have
gone through multiple CCU, TCU, RCU and FCU boards swaps and ECL level
changes to no avail. I was curious if anyone else had seen (and
hopefully fixed) a similar error.


_____________________________________________________________
_______
Error string:
"Error in copmem:- cannot send command to coprocessor (timeout flag
=1
). Check for bad RCU/TCU or crashed stadtd on CCU.""
(say OK to this error message and we get)
Rcustart cannot start RCU
(say OK to this meassge and we get)
Cannot start acquisition: timeout in TCU
_____________________________________________________________
_______





Thanks



Todd M. Alam
Sandia National Laboratories
NMR Facility, Org 1816, MS 0886
Dept of Electronic Materials and Nanostructured Materials
(505) 844-1225
(505) 844-2974 (FAX)
tmalam_at_sandia.gov
http://www.sandia.gov/materials/sciences/nmr_lab/