31P-103Rh NMR

From: Charles G. Fry <fry_at_chem.wisc.edu>
Date: Thu, 10 Aug 2000 17:44:23 -0500

Dear AMMRLs:

Having fun doing 103Rh for the first time. Want to observe in
samples that are not so soluble, so we're using 31P excite/detect
in HMQC and HSQC type experiments. No problems see the 103Rh
chem shifts, but are having two problems I'm hoping this group
can assist with:

1. We want to see 31P coupling in the 103Rh dimension, so we can
   count # phosphorous bound to 103Rh. 31P would almost certainly
   be sufficient, but the Prof. would like to get the hetero-coupled
   multiplet from the 103Rh F1 traces. So interesting problem:
   how to get a heteronuclear J to show in the t1/F1 dimension.
   I'm trying a couple obvious ideas, but would like to know of
   publications/experiments that tackle this problem.

2. Probe is a 1H observe + lock on the inner coil, and 31P fixed +
   X (nominal 13C-109Ag tunable) on the outer coil. We modified the
   probe so X could tune down to 15.8MHz for 103Rh. We're
   delivering 400 watts and have Q=52 at 103Rh. I find pw90=150us.
   Ouch... Even though 103Rh is on the outer coil on this newer
   style Nalorac probe, I definitely thought we would do better.
   Had our engineer look at the probe in some detail on a spectrum
   analyzer, but he insists everything is ok (e.g., no close-by
   resonances).

   Any of you that work down at these frequencies: I know the pw90
   is coil/probe dependent, but what pw90's are you getting?
   Any suggestions what we should check? The probe does not show
   symptoms that would normally show up for arcing.

Thanks,
Charlie

----------------------------------------------------------