Re: F19-C13 ghmqc on Inova

From: Wei Li 448-7532 <wli_at_utmem.edu>
Date: Tue, 18 Sep 2001 09:38:23 -0500

Hello,

I believe your second channel is H1-only, check the configuration of the
hardware. The first channel should be a full band. When you have dn='H1',
the 2nd channel (H1 only) will be used as decoupler and the 1st channel
(full band) is used as transmitter, therefore you don't have any problem to
get F19 frequency out. When you set dn to anything except 'H1', the first
channel will automatically become the decoupler channel since it is a full
band, while the 2nd channel is the transmitter, and there is no surprise
that H1 only synthesizer won't get F19 frequency out.

Your 3rd channel should be a low band channel, and you should use this as
the decoupling channel while you use channel one as xmitter.

Good luck
-----------------------------------------
Wei Li, Ph.D
Dept. of Pharmaceutical Sciences
University of Tennessee HSC
847 Monroe, Room 327
Memphis, TN 38163

Phone 901-448-7532 (Office) Fax 901-448-6828 (Office)
Phone 901-753-6978 (H)
----- Original Message -----
From: "Ion Ghiviriga" <ion_at_chem.ufl.edu>
To: <ammrl_at_chemnmr.colorado.edu>
Sent: Monday, September 17, 2001 10:35 AM
Subject: F19-C13 ghmqc on Inova


> Dear AMMRL Members:
>
> I am trying to set up a F19-C13 ghmqc experiment on an Inova with 3
> rf channels. The probe is tuned to F19 and I get a good s/n in a F19
> spectrum, as long as dn='H1'. When I turn dn=C13 (or dn='F19',
> homo='y)' all I get is noise. In both cases, dm='n'.
> Any idea why this happens?
> Thank you,
> Ion
>
> Ion Ghiviriga
> Coordinator of NMR Services
> Department of Chemistry, University of Florida
>
> 393 Leigh Hall
> PO Box 117200
> Gainesville, FL 32611-7200
> Phone: 352-846-3001
> e-mail: ion_at_chem.ufl.edu
> http://www.chem.ufl.edu/NMR/Ion/ion.htm
>
>
>
>
>
Received on Tue Sep 18 2001 - 10:25:05 MST