Re: LINUX communication with Bruker Aspect-3000

From: Gary Juneau <nmr_at_decaturcenter.com>
Date: Thu, 23 Aug 2001 12:20:35 -0500

Dear Dr. Bangerter,

I'm trying to connect an old Varian VXR to a PC. I have the communication program from
Varian, which supposedly allows me to xfer files from the instrument's computer to a
UNIX machine, via RS232 ports. However, I still don't have a UNIX communications
program, e.g, cu. I'm wondering if LINUX had some sort of communications program in
one of it's packages. Could I use LINUX to complete the connection?

Thanks! Gary Juneau

Ben Bangerter wrote:

> Yes, ISA slots in PCs are rapidly becoming a thing of the past. If the Metrabyte
> board you are familiar with is the one which is emulated by the Measurement
> Computing (formerly Computer Boards) CIO-DIO24, which is the case with the original
> ZZNET scheme (and NMRNET, subsequently), this is your lucky day! Measurement
> Computing offers a PCI card, the PCI-DIO24 (cost $89), which also incorporates the
> 82C55 PIO chip. This board has the same 37 pin D connector and is "fully
> compatible" with the ISA board. Go to www.measurementcomputing.com and have a
> look. I haven't used this PCI board, but do use the ISA version.
>
> We (and many others) have been using NMRNET and a Windows PC to transfer data from
> AM and AC systems and processing with NUTS or other third party software. This I
> have found works very well, and certainly enhances the utility of those older A2000
> based consoles. (Just don't get me started on BRUKNET, LIGHTNET, the COMM box...)
>
> Ben Bangerter
>
> *************************
>
> Neil Jacobsen wrote:
>
> > I have been working for some time on a way to run XWinNMR on a Linux
> > machine and connect it in a convenient way to our Bruker AM-250. I am
> > using a program called SENDZZ which Tom Pratum developed from the original
> > ZZNET program. SENDZZ allows you to enter at the Bruker AM-250 keyboard:
> >
> > PASC SENDZZ <filename>
> >
> > and the file is sent to a PC running the corresponding C program. I have
> > been working with a friend who is a software engineer specializing in
> > communication and networking, and he has modified the C program so it will
> > run on a modern PC and communicate with the Aspect 3000. Now we are
> > preparing to buy a new PC with LINUX operating system and get it to talk to
> > the Aspect. The problem right now is that we can't find a new PC which has
> > a slot for the old Metrabyte I/O card. So we are looking for a new card
> > which will do the same thing.
> >
> > I'm wondering how many people out there have tried to do the same thing?
> > Maybe this problem has already been solved. My goal is to have data
> > transferred after acquisition to the PC in a fast and convenient way, and
> > then processed on the Linux box using XWinNMR. This would be a cheaper
> > alternative to the TechMag upgrade but would still offer convenience in
> > data processing, storage and backup. The cost, including software, would
> > be about $5K compare to the $40K or more for TechMag. Any advice,
> > suggestions, and anecdotes would be welcome.
> >
> > Neil
> >
> >
> > Neil E. Jacobsen, Ph.D.
> > NMR Facility Manager
> > Department of Chemistry
> > 119 Old Chemistry
> > 1306 E. University
> > University of Arizona
> > Tucson, AZ 85721
> > 520-621-8146
> > FAX 520-621-8407
Received on Thu Aug 23 2001 - 12:23:31 MST