Re: printers & Jeol EXs

Steve Philson (philson@nmr.chem.umn.edu)
Fri, 16 Apr 1999 10:59:10 -0500

On Thu, Apr 15, 1999 at 06:42:48PM +0100, Phil Dennison wrote:
>
> I'm replacing the HP plotters on my spectrometers with A4 laser printers.
> This is fine on the Bruker spectrometers, as drivers are available, and
> the older machines print happily via serial to parallel converters.
>
> However, I also have a Jeol EX270 for which Jeol UK tell me there is no
> replacement option for the HP 7550a plotter.
>
> Has anyone come up with a solution for this?

I had a similar problem with plotting from Bruker AC consoles, and
also from an EPR machine. Fortunately, the handshaking only took
place at the start of the plot, so it was fairly simple to eavesdrop
on the serial connection, figure out just what the console wanted to
see, and then stick in a modest PC to emulate the plotter, receive
the data, and send it off to whatever kind of printer I felt like
hooking up to the PC. It sounds like your situation may be more
complicated, if there is periodic communication with the plotter,
but you could still probably do something like the same thing.

I used a neat little piece of software called serialtest made by
Frontline Test Equipment, Inc.; the company has newer versions with
more capabilities and simpler requirements on the eavesdropping
computer. Their web site is http://www.fte.com. There are probably
free ways to accomplish the same thing, but I haven't looked into
them; the software was not very expensive.

Once the requirements are known (and it is possible that someone
with Jeol could give you the information without the need for
eavesdropping), it is fairly straightforward to write a little
program (I used perl, on a FreeBSD or Linux OS) to do whatever
emulation you need.

Good luck.

-- 
Steve Philson           philson@nmr.chem.umn.edu
Director NMR Lab        612-626-0297
Chemistry Dept.         University of Minnesota