We use a direct drive Atlas Copco LE-7 compressor
mounted on a horizontal 80 gallon tank which the
vendor's literature terms an Atlas Copco model
D5-80H. Atlas Copco Airpower n.v. is a Belgium
firm which has designed a very nice low oil in
air entrainment compressor. Our compressor is
set to run between 90 and 120 PSI, and requires
almost no maintenance.
This model costs around $2200.
In addition, in our air system, we placed a Balston
"D" series coalescing filter {model 92-812 holder
with the 100-12-DX filter} in line before an Arrow
A25 refrigerated air dryer and a Balston "B" series
coalescing filter {Model A915A-BX with 100-18-BX
filter} after the air dryer. With three spectrometers
simultaneously ejecting, we have not noticed a pressure
droop.
The refrigerated air dryer, and two Balston filter
units combine to add about another $1200 to the total
cost.
I hope this helps.
Bruce D. Ray
IUPUI Physics Dept.
NMR Center