Bruker AC-200 SCM Field value changing

From: Lawrence Byrnes <numare_at_ptd.net>
Date: Fri, 22 Apr 2005 09:00:59 -0400

Hello Spinlanders,

Have observed following with respect to 'Field' value on console system
control microprocessor with same sample in CDCl3:

04 March 2005, Field = - 4611
06 " " " -4611
14 " " " -4692
23 " " " -4820
30 " " " -4899
07 April " " -4978
17 " " " -5112
18 " " " -5119
19 " " " -5128
20 " " " -5162
21 " " " -5165
22 " " " -5158

Above values are the values selected by the 'Autoshim' procedure after
shimming sample for optimal signal.

System was last pumped and energized 14 January 2005. System was last
cryoshimmed 17 February 2005.

Have any of you folks made similar observations? If so, what problem(s) lie
ahead for our NMR system and what are the preventative/restorative
maintenance procedures that are recommended at present time?

Will post summary of helpful responses.

Thank you and kind regards,

Lawrence


Lawrence Byrnes,

Lab Manager and
Research Director

Numare Spectralab and
The Schoenenberg Institute of Molecular Spectroscopy

Hidden Meadow Farm
3551 Winding Road,
Kintnersville, PA, 18930

Voice: 610-346-8660
Fax: 610-346-7571
E-mail: numare_at_eclipse.net
Web: http://www.eclipse.net/~numare/

      IMPORTANT NOTICE! PLEASE READ!

THIS EMAIL, AND ANY FILES TRANSMITTED WITH IT, ARE CONFIDENTIAL AND
INTENDED SOLELY FOR THE USE OF THE INDIVIDUAL OR ENTITY TO WHOM THEY
ARE ADDRESSED. IF YOU ARE NOT THE INTENDED RECIPIENT, YOU ARE NOT
PERMITTED TO DISTRIBUTE OR USE THIS MESSAGE OR ANY OF ITS ATTACHMENTS
IN ANY WAY. WE ALSO REQUEST THAT YOU ADVISE THE SENDER OF THE INCORRECT
ADDRESSING. THIS NOTICE ALSO CONFIRMS THAT THIS EMAIL MESSAGE HAS BEEN
VIRUS SCANNED AND THAT, ALTHOUGH NO COMPUTER VIRUSES WERE DETECTED,
NUMARE SPECTRALAB INCORPORATED ACCEPTS NO LIABILITY FOR ANY CONSEQUENTIAL
DAMAGE RESULTING FROM EMAIL CONTAINING ANY COMPUTER VIRUSES.
Received on Fri Apr 22 2005 - 17:45:26 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 10 2023 - 17:33:13 MST